Ultrasonic Flaw Detector Novotest Ud2301 (mini)
Product Specification:
| Brand | PRO Engineers Novotest | 
| Display Type | high resolution (320*480 pixels) | 
| Weight | 250 g (without batteries) | 
| Frequency | from 1 to 10,0 MHz | 
| Dimension | 80x162x38 mm | 
| Minimum Order Quantity | 1 Piece | 
Product Description
The picture on the display of the device can be rotated in 4 position, that allows to operate the device in any position, making it customisable for both: lefties and righties users, as well as positioning the Instrument in portrait and landscape orientations.
The use of the ultrasonic flaw detector:
to test the quality of welds and base metal of products;
to detect corrosion, cracks, delaminations and other internal defects;
to determine coordinates and parameter estimation of defects such as discontinuity, cracks and uniformity of material in products from metal, plastic, composites and other materials;
to measure the thickness of products.
The Flaw detector is equipped with 3 standard type AA rechargeable batteries. The device can also operate with non-rechargeable batteries of the same type. Using the batteries of standard type allows autonomous operation of the device in almost any conditions – if there is no possibility to charge batteries, it is possible to have spare batteries, or buy new batteries at any store. The Flaw detector allows user to solve a wide range of tasks – from thickness measurement of thin products to large-sized casting flaw detection. Having sealed housing with rubber protective strips makes Ultrasonic flaw detector UD2301-mini ideal for using in workshop and field conditions with high humidity, dust, etc. The device has frost-resistant display that allows to use the device at any season and in any climatic zone of the Earth.
| perating frequency range | from 1 to 10,0 MHz | 
| Range of measured time intervals (duration of scanning) | from 6 to 1000 µs | 
| Velocity range | 1000 – 9999 m / s | 
| Error of measurement of time intervals | not exceed ± 0,025 µs | 
| Maximum permissible error of measurement of the amplitudes of the signals at the receiver input in the range from 0 to 110 dB | not exceed ± 0,5 dB | 
| Testing gain range | 125 dB | 
| Averaging over the quantity of starts | from 1 to 16 | 
| Range of variation of temporal sensitivity adjustment (TVG) | 40 dB | 
| Number of control points TVG | 15 | 
| Duration of the excitation pulse to the load | from 0,0 to 0,5 µs | 
| Amplitude of the excitation pulse to the 50 Ohm load, not less | 100, 200, 300 | 
| Operating frequency range of the receiver on the level -3 dB | from 1 to 10 MHz | 
| Deviation of the amplitudes of input signals in the range from 10 to 100% of the screen height not more than | 1 dB | 
| Scanning | from 1 to 1000 µs | 
| Delay of scanning | from 0 to 2000 µs | 
| Range of measurement of time intervals | from 0 to 1000 µs | 
| Setting a delay in the prism of the probe | from 0 to 15 µs | 
| Automatic signaling of defects (AFS) | dual-gate | 
| Setting range of AFS gates | from 0 to 2000 µs | 
| Adjusting thresholds of AFS gates | from 0 to 100 % of screen height | 
| Detection of signals | positive half-wave, radio mode | 
| Dimensions (W*H*L) | 80*162*38 mm | 
| Weight, not more | 250 g (without batteries) | 
| Mean time between failures | not less than – 3000 hours | 
Additional Information :
| Delivery Time | 7 days | 
| Port Of Dispatch | MUMBAI | 
| Production Capacity | 100 | 
| Packaging Details | Ultrasonic flaw detector UT-probes – 2 pcs Cable Lemo-Lemo – 1 pc AA batteries – 3 pcs Charger USB cable for PC connection Operating manual Case | 
| Payment Terms | T/T (Bank Transfer)/Cheque | 